An Ultra-Sensitive Label-Free Immunosensor Toward Alpha-Fetoprotein Detection Based on Three–dimensional Ordered IrOx Inverse Opals

Kuang Sheng,Wei Liu,Lin Xu,Yandong Jiang,Xinran Zhang,Biao Dong,Geyu Lu,Hongwei Song
DOI: https://doi.org/10.1016/j.snb.2017.07.116
IF: 9.221
2017-01-01
Sensors and Actuators B Chemical
Abstract:Three-dimensional ordered macroporous iridium oxides (3DOM IrOx 0 <= x <= 2), were synthesized through a simple self-assembly template method and developed as an electrode material for lable-free immunosensor of cancer biomarker alpha-fetoprotein (AFP). The 3DOM IrOx was well modified on flexible fluorine-doped tin oxide (FTO) electrode and such FTO electrodes based 3DOM IrOx immnuosensor enhanced the adsorption of electrolytes, the diffusion of reactants, and also could act as a rapid electron transfer channel benefited both from the advantage of 3DOM architecture and the nature of IrOx itself. These combined characteristics greatly increased the sensitivity, shortened the response time, reduced the interference, promoted the stability and moreover, demonstrated good accuracy to detect AFP inhuman serum. The optimum structure and sensing properties of immnuosensor were obtained based on 3DOM IrOx annealed at 500 degrees C. It allows detecting AFP over a wide concentration range from 1 to 250 ng/mL with a detection limit of 0.3 ng/mL, and a higher slope of the calibration equation. The manufactured immunosensor was also used to determine AFP in human serum with satisfactory results. This kind of 3DOM IrOx modified and FTO-based immunosensor has great application potential for the AFP electrochemical detection. (C) 2017 Published by Elsevier B.V.