Combining Field-Amplified Sample Stacking with Moving Reaction Boundary Electrophoresis on A Paper Chip for the Preconcentration and Separation of Metal Ions

Liangfei Ouyang,Qian Liu,Heng Liang
DOI: https://doi.org/10.1002/jssc.201600804
IF: 3.614
2017-01-01
Journal of Separation Science
Abstract:A common drawback of paper-based separation devices is their poor detection limit. In this study, we combined field-amplified sample stacking with moving reaction boundary electrophoresis on a paper chip with six array channels for the parallel separation and concentration of multiple samples. With a new hyphenated technique, the brown I2 from the Fe3+ /I- oxidation-reduction reaction emerged near the boundary between the dilute ethylene diamine tetraacetic acid and potassium iodide and highly concentrated KCl solutions. For the separation and concentration of three components, Cr3+ , Cu2+ , and Fe3+ , the Fe3+ detection limit was improved at least 266-fold by comparing the hyphenated technique with moving reaction boundary electrophoresis. The detection limit of Fe3+ was found to be as low as 0.34 ng (20 μM) on the paper chip. We also demonstrated the analysis of a real sample of four metal ions, with detection limits as follows: 0.16 μg Cr3+ , 1.5 μg Ni2+ , 0.64 μg Cu2+ , and 1.5 μg Co2+ . The synergy of field-amplified sample stacking and moving reaction boundary electrophoresis in the micron paper-based array channels dramatically improved the detection limit and throughput of paper-based electrophoresis.