A Novel Magnetism‐assisted Electrochemical Immunosensor with Sub‐Picomolar Sensitivity

Pengli Li,Hongfang Zhang
DOI: https://doi.org/10.1002/jccs.201600120
2016-01-01
Journal of the Chinese Chemical Society
Abstract:A novel electrochemical immunosensor based on a magnetic glassy carbon electrode (MGCE) was developed for the quantitative determination of human immunoglobulin G (IgG). The immunosensing interface was fabricated by initially depositing silver nanoparticles on the MGCE surface and then immobilizing anti-human IgG antibodies via the magnetic force between MGCE and Fe3O4 nanoparticles. The antibodies were covalently bonded to the amine-functionalized Fe3O4 nanoparticles. Under optimal conditions, the magnetism-assisted immunosensor exhibited a wide linear range from 0.1 pg/mL to 1.0 mu g/mL with the detection limit of 0.05 pg/mL. Furthermore, the immunosensor displayed the advantages of good reproducibility and satisfactory stability.
What problem does this paper attempt to address?