A Sandwich-Type Electrochemical Immunosensor Based on the Biotin- Streptavidin-Biotin Structure for Detection of Human Immunoglobulin G

Yueyun Li,Yihe Zhang,Liping Jiang,Paul K. Chu,Yunhui Dong,Qin Wei
DOI: https://doi.org/10.1038/srep22694
IF: 4.6
2016-01-01
Scientific Reports
Abstract:A sandwich-type immunosensor is designed and fabricated to detect the human immunoglobulin G (HIgG) using polyaniline and tin dioxide functionalized graphene (GS-SnO2-PAN) as the platform and biotin-functionalized amination magnetic nanoparticles composite (B-Fe3O4@APTES) as the label. GS-SnO2-PAN is used as the sensing agent to capture the primary anti-HIgG (Ab1) and SnO2 reduces the stack of GS. The B-Fe3O4@APTES with a large surface area and excellent biocompatibility captures second antibody (Ab2) efficiently based on the highly selective recognition of streptavidin to biotinylated antibody. The B-Fe3O4@APTES has better electro-catalytic activity in the reduction of hydrogen peroxide (H2O2) and the "biotin-streptavidin-biotin" (B-SA-B) strategy leads to signal amplification. Under optimal conditions, the immunosensor has a wide sensitivity range from 1 pg/L to 10 ng/L and low detection limit of 0.33 pg/L (S/N = 3) for HIgG. The immunosensor has high sensitivity, fast assay rate, as well as good reproducibility, specificity, and stability especially in the quantitative detection of biomolecules in serum samples.