A Method for Determining Thermal Conductivity of Tissues in Temperature Range 233 ~ 293K

张海峰,程曙霞,何立群,张爱丽,郑瑛,高大勇
DOI: https://doi.org/10.3969/j.issn.0253-2778.2003.02.011
2003-01-01
Journal of University of Science and Technology of China
Abstract:A novel method using a tiny thermistor as a spherical heat source has been developed to determine thermal conductivity of biomaterials over a wide temperature range. Based on steady-state spherical heat transfer in an infinite homogeneous medium, thermal conductivity of the measured medium can be determined from the power applied and the temperature rise of the self-heated thermistor. Compared with recommended value, maximum measurement error of thermal conductivity for some standard samples is 5.1% in the temperature range 233~313K. Using the developed technique, thermal conductivities of rabbit liver, kidney, heart, and carotid artery were determined in the temperature range 233~293K; the results show that there exists a significant difference between pure ice and frozen tissues below zero temperature.