Evaluation of Fetal Heart Using Fetal Heart Quantification (fetalHQ) Technique in an Indian Cohort

Shreyasi Sharma,Mar Bennasar,Manisha Yadav
DOI: https://doi.org/10.1002/jcu.23859
2024-10-13
Journal of Clinical Ultrasound
Abstract:FetalHQ reliably estimates global cardiac morphometry but offers variable results for global cardiac function. Regional transverse contractility, shown by fractional shortening, poorly correlates for the right ventricle and moderately to poorly for the left. Despite this, FetalHQ is a promising tool needing technical advancements and validation for improved accuracy. Objective To evaluate the intra‐ and interobserver reproducibility of parameters used to determine cardiac morphometry and deformation using FetalHQ in Indian population. Methodology Two operators blinded to each other performed FetalHQ analysis independently on 35 normal fetuses of gestational age between 18 + 0 and 37 + 0 weeks and days. Intra‐ and interobserver correlation coefficient and limits of agreement were ascertained for parameters obtained from HQ analysis. Results Excellent reproducibility (ICC > 0.9) was observed for global morphometric parameters. Ejection fraction, stroke volume, stroke volume/kg demonstrated excellent reliability (ICC > 0.9), cardiac output, fractional area change showed a lower correlation (ICC