On-line Quantitative Analysis of Heavy Metals in Water Based on Laser-Induced Breakdown Spectroscopy

N. J. Zhao,D. S. Meng,Y. Jia,M. J. Ma,L. Fang,J. G. Liu,W. Q. Liu
DOI: https://doi.org/10.1364/oe.27.00a495
IF: 3.8
2019-01-01
Optics Express
Abstract:Heavy metal pollution from industrial wastewater is an important source. A method for heavy metals determination in industrial wastewater based on laser-induced breakdown spectroscopy (LIBS) technique was studied and the on-line monitoring system that used automatic graphite enrichment and spatial plasma confinement detection was developed and field demonstrated. The limits of detection (LOD) of heavy metal elements (Cd, Cr, Cu, Ni, Pb, Zn) could reach several μg/L. In Tongling, the on-line heavy metal monitor was field demonstrated. The calibration curves of copper and zinc were built on site, and then on-line monitoring was conducted. The measurement results of this monitor were compared with ICP-OES and had a good correlation. The results showed that the heavy metal monitor could be used for on-line detection of heavy metals in wastewater and had a good reliability.
What problem does this paper attempt to address?