A Minimally Invasive Hollow Microneedle with a Cladding Structure: Ultra-Thin but Strong, Batch Manufacturable

Jiaming Chen,Ping Cheng,Cecilia Sun,Yipin Wang,Xiaojing Zhang,Zhuoqing Yang,Guifu Ding
DOI: https://doi.org/10.1109/tbme.2019.2906571
2019-01-01
Abstract:OBJECTIVE:A minimally invasive hollow in-plane microneedle with a cladding structure is designed to improve the mechanical strength.METHODS:The traditional weak stack structure has been changed into a cladding structure, and the effectiveness has been validated through finite element analysis. The prototypes of the microneedles were batch manufactured by the integrated micromachining process with no need to assemble.RESULTS:Compared to our previously reported microneedle with the weak stack structure, the cladding microneedle in this paper shows 263% improvement in bonding strength (6.4±0.30 N) and 36.5% improvement in buckling strength (2.8±0.07 N). In addition, the fabricated microneedle will not fail during insertion into the fresh and dehydrated pig skin with a satisfying safety factor (1.55).CONCLUSION:A novel structure of hollow microneedle was developed and fabricated by microfabrication technology. The improvement in mechanical strength is obvious.SIGNIFICANCE:The microneedle has great mechanical property and good potential for wider applications in human skin.
What problem does this paper attempt to address?