Screen-printed Biosensor System Based on Sequential Injection Analysis Method

Fang ZHAO,Qiang CAI,Xiao-He YANG,Xiang-Qiang CHEN,Fang-Yi PENG,Hui-Min PENG
DOI: https://doi.org/10.3724/SP.J.1096.2011.00511
IF: 1.193
2011-01-01
Chinese Journal of Analytical Chemistry
Abstract:The reaction system of horseradish peroxidase (HRP) catalyzing o-phenylenediamine (OPD)-H(2)O(2) using biosensor coupled with sequential injection analysis (SIA) was studied. The biosensor was screen printed electrode immobilized with bovine serum albumin (BSA) and glutaraldehyde. After HRP sensor on the SIA system, the experiment of sensor's dynamic response for H(2)O(2)-OPD was studied, including a series of optimization, such as the buffer pH and concentration, pump flow rate, sample volume. The result showed that under the optimum conditions of the experiment, including flow rate of 153. 9 mu L/min, sample volume of 100 mu L and the buffer of pH 6. 5(0. 1 mol/L), the HRP biosensors has a linear response in the range of 0. 1 - 1. 0 mmol/L for concentration of H(2)O(2). The current response curve increased with the increase of the concentration of H(2)O(2). In addition, the biosensor had the advantages of repeatability with RSD of 3. 8%. The recovery ration for OPD in the real sample was more than 60%.