Ceramic-on-ceramic total hip arthroplasty:Pitfalls and strategy

CHEN Yun-su,SHAO Jun-jie,s H
DOI: https://doi.org/10.3969/j.issn.1673-7083.2009.02.007
2009-01-01
International Journal of Orthopaedics
Abstract:Objective To analyze the pitfalls in ceramic-on-ceramic total hip arthroplasy(THA),and find the way to avoid the complications.Methods From September 2001 to December 2007,we operated on 1019 hips with ceramic-on-ceramic total hip prosthesis in our department.The patients included male 585 and female 434,mean age 59.5(range 39-76).Prosthesis used includ 362 Trident cups & Secur-fit Max stems(Stryker),544 EP-fit Plus cups & SL-Plus stems(Plus Orthopedics) and 113 SPH cups & C2 stems(Lima,sandwich type).Preoperative diagnosis include femoral head avascular necrosis(314 cases), developmental dysplasia(dislocation) of the hip(268 cases),osteoarthritis(194 cases),ankylosing spondylitis(74 cases), rheumatoid arthritis(107 cases),and femoral neck fracture(GardenⅣ,62 cases).Results Cup position was adjusted in 34 cases because of hip instability during intra-operative test.Ceramic liners were replanted in 19 cases.There were 3 ceramic liner chipping and leg length shortening more than 0.5 cm in 31 cases.Conclusion To avoid complications,special technique should be learned while using ceramic-to-ceramic total hip prosthesis.
What problem does this paper attempt to address?