An Atypical Adverse Local Tissue Reaction After Ceramic-on-Ceramic Primary Total Hip Arthroplasty

Kamran Movassaghi,Arpan Patel,Ira Miller,Brett R Levine
DOI: https://doi.org/10.1016/j.artd.2022.01.025
2022-02-28
Abstract:Adverse local tissue reaction is an uncommon but frequently described complication after total hip arthroplasty (THA). It is most often associated with metal-on-metal hips and less frequently with metal-on-polyethylene implants as part of a mechanically assisted crevice corrosion process. In this report, we describe a rare case of an atypical adverse local tissue reaction in a patient with a ceramic-on-ceramic THA. Abrasive backside liner wear from a prominent screw head, failure of the liner locking mechanism, and liner fragmentation secondary to component-component impingement created an atypical mass and fluid collection leading to THA failure. This case demonstrates the importance of appropriate cup-liner positioning, thorough workup of pain after THA, and the ability of ceramic debris to cause an associated, atypical adverse local tissue reaction.
What problem does this paper attempt to address?