Studies of frequency doubling perimetry in detection of glaucomatous visual field loss

LU Chen-Ming,XU Cheng-Hui,ZHU Zhao-Rong
2006-01-01
Abstract:Objective To evaluate the ability of frequency doubling perimetry in detecting glaucomatous visual field loss,and to ascertain whether FDP could provide results comparable with those of automated OCTOPUS 101 perimetry.Methods The visual fields of the participants were assessed by OCTOPUS program G2-TOP and FDP program C-20-5 screening.The participants include 23 normal,20 early glaucoma,35 moderate-advanced glaucoma,11 high ocular hypertension and 13 suspected glaucoma.Only 1 eye of each participants was selected randomly except the glaucoma group,in which the eye with more severe visual field loss was selected.Results For the FDP program C-20-5 screening,area under the ROC curve is 0.925(sensitivity 85%,specificity 91%)in detecting glaucomatous visual field loss.The pearson coefficient between FDP results and OCTOPUS indices-mean defect(MD)and loss variance(LV)are 0.702 and 0.429(P<0.001),respectively.The time cost by FDP and OCTOPUS101 perimetry are 1.00 min and 2.33 min,respectively.A significant difference was found between the two visual field tests(P<0.001).Conclusion Frequency doubling perimetry demonstrates high sensitivity and specificity in the detection of glaucomatous visual field loss.A significant correlation was fould between FDP and OCTOPUS indices.The FDP is a faster way to analyze the visual field than OCTOPUS101 perimetry.All these foundings make a big population screening possible.