Application of 125I in Micropipettor Check

QIAN Xiao-yu,GAO Yun-chao,YANG Ya-dong,LU Han-kui
DOI: https://doi.org/10.3969/j.issn.1000-7512.2006.03.014
2006-01-01
Abstract:Gravimetric test is one of authoritative methods for the measurement and validation of the accuracy and precision of micropipettors.The feasibility of radioactive measurement in the check of micropipettors is studied.There are no significant difference in radioactive count rates of ~(125)I aspirated between 2 qualified micropipettors,but significant difference are recorded between qualified and unqualified micropipettors.So ~(125)I radiometry agrees with gravimetric tests,and can be used as a laboratory constant test of micropipettors.
What problem does this paper attempt to address?