[Accuracy of Injection Dose of Amyloid PET Agent Using Radiopharmaceutical Activity Suppliers]

Yukito Maeda,Keiichi Matsumoto,Yasuhiko Ikari,Go Akamatsu,Keiji Shimizu,Keisuke Tsuda
DOI: https://doi.org/10.6009/jjrt.2024-1423
2024-02-20
Abstract:Purpose: This study aimed to identify disposable items with low amyloid positron emission tomography (PET) agent radioactivity adsorption for accurate injections using a radiopharmaceutical activity supplier. Methods: First, we investigated disposable items currently used for amyloid PET agent injection. Next, we measured the residual radioactivity rates of amyloid PET agents on three-way stopcocks, extension tubes, butterfly needles, and indwelling needles to identify disposable items with low radioactivity adsorption. Finally, we evaluated the accuracy of amyloid PET agent injection using the selected disposable items and a radiopharmaceutical activity supplier. Results: The polybutadiene extension tube exhibited a significantly lower residual activity rate than that of the polyvinyl chloride extension tube. Similarly, the indwelling needles showed significantly lower residual activity rate than that of butterfly needles. The dose indicated by a radiopharmaceutical activity supplier was 184.1 MBq, while the dose calibrator measured the radioactivity which flowed into the vial as 170.2 MBq, resulting in an administration accuracy of 8.2%. Conclusion: To ensure accurate amyloid PET agent injections, we recommend using polybutadiene extension tubes and indwelling needles due to their lower radioactivity adsorption.
What problem does this paper attempt to address?