0491 VALIDATION OF THE NOX-T3 PORTABLE MONITOR FOR DIAGNOSIS OF OBSTRUCTIVE SLEEP APNEA IN CHINESE ADULTS
Liang Xu,Fei Han,Brendan T Keenan,Elizabeth Kneeland-Szanto,Hui-chan Yan,Xiaojing Dong,Yulan Chang,Long Zhao,X Zhang,Juan-Juan LI,Allan I Pack,S Kuna
DOI: https://doi.org/10.1093/sleepj/zsx050.490
IF: 6.313
2017-01-01
SLEEP
Abstract:To evaluate the performance of a portable monitor (Nox-T3, Nox Medical Inc. Reykjavik, Iceland) used to diagnose obstructive sleep apnea in Chinese adults. Eighty Chinese adults (mean age 47.6 ± 14.0 years, 77.5% males, BMI 27.5 ± 5.4 kg/m2) performed an overnight, unattended home sleep test (HST) with the Nox-T3 portable monitor followed by an overnight in-laboratory polysomnogram with simultaneous portable monitor recording. The portable monitor recordings were scored using automated analysis and then manually edited using different criteria for scoring hypopneas. Polysomnograms were scored based on recommended guidelines. When scoring of hypopneas required a ≥4% oxygen desaturation event, the mean apnea-hypopnea index (AHI) was 24.4 ± 20.8 events/hour on HST, 28.0 ± 22.9 events/hour on in-lab portable monitor recording, and 28.6 ± 23.9 events/hour on PSG (p<0.0001). Bland Altman analysis of AHI on PSG versus HST showed a mean difference (95% confidence interval) of -4.64 (-7.15, -2.13); limits of agreement (equal to ±2SD) was -26.62 to 17.35 events/hour. Based on a threshold of AHI≥5 events/hour, HST had 95% sensitivity, 69% specificity, 94% positive predictive value, and 75% negative predictive value compared to PSG gold-standard. Using an AHI≥15 events/hour, HST had 93% sensitivity, 85% specificity, 89% positive predictive value, and 91% negative predictive value. Closer agreements were present when comparing the simultaneous recordings. Similar results were obtained using different scoring criteria for hypopneas. Despite known differences between HST and PSG, the results show close agreement between the two diagnostic tests in Chinese adults, especially when controlling for night-to-night variability and changes in sleeping environment. STK was supported by NIH HL094307, FH was supported by research grants from the Ministry of Science and Technology (2014DFA31500) and Beijing Municipal 1 Science & Technology Commission No. Z161100002616012 Correspondence: Fang Han(hanfang1@hotmail.com).