Preliminary results on visualization of radiofrequency ablation lesions using real-time ultrasound elastography with the bovine liver ex-vivo

LI Jia,HU Bing,HU Bin,CHEN Lei
DOI: https://doi.org/10.3969/j.issn.1672-6448.2009.04.007
2009-01-01
Abstract:Objective To evaluate the reliability and feasibility of real-time ultrasound elastography for estimating the RFA-lesion boundaries in liver.Methods Bovine livers were chosen for this study.Twenty-four RFA were performed on bovine livers with different depth including 1 cm,3 cm and 5 cm.Each group had eight RFA-lesions.Gray-scale ultrasound and elastography were performed after the RFA.The imaging planes were perpendicular and parallel to the axis of the RF electrode.All the ultrasound and elastography images were stored in computer.Then the measurements of all the RFA-lesions were done.Measurements of the sections representing the same image planes used for elastography were taken at pathologic examination and compared with the measurements obtained from the gray-scale ultrasound images and elastograms.Results The average volumes of gray-scale ultrasound images,elastograms and pathologic measurements of 1cm group were 514.8±63.4 mm3,748.1±98.1 mm3,766.3±43.2 mm3 respectively.The average volumes of 3 cm group were 498.1±154.9 mm3,694.4±48.1 mm3,707.5±25.0 mm3 respec-tively.The RFA-lesions could not be evaluated by grey-scale ultrasound correctly(P<0.05).RFA-lesion boundaries in the elastograms were more clearer in 1 cm group and 3 cm group.We found a statistically significant correlation between elastography and pathology measurements(r=0.6064,P<0.05).The RFA-lesions could not be estimated by lastography in case the depth increased to 5 cm.Conclusions Real-time ultrasound elastography may prove to be reliable for estimating the boundaries of the RFA-lesions and be useful for monitoring during the RFA procedure.
What problem does this paper attempt to address?