刘素芹,戴高鹏,王启会
DOI: https://doi.org/10.3969/j.issn.1000-9787.2009.04.021
2009-01-01
Abstract:8 independent SPQC sensors is parallel connected to form a muti-channels detecting unit.Electrodes are immersed into culture medium to detect conductivity change of solution.The cell constant k is determined as 0.05m,and the sensitivity interval of the device is from 550 to 600μs.To maintain sensitivity of SPQC sensor,a novel culture medium(a mino acid broth) AaB is developed.FDT is obtained veraciously by using fitting-differentiating method.Common pathogens S.aureus and S.dysenteriae are determined with the automated detecting device and methods mentioned above.Results show that the calibration curves of FDT against density of bacteria show a linear correlation coefficient(R0.99) over the range of 10~106 CFU·mL-1.Detection results are all inside the 95% confidence interval of standard pour plate counting method.The reproducibility is also reviewed,and the device is still stable and sensitive.