Spatial and Temporal Amperometric Monitoring Dopamine Release from Single PC12 Cells with Carbon Fiber Nanoelectrode

黄卫华,张丽瑶,程伟,庞代文,王宗礼,程介克
DOI: https://doi.org/10.3321/j.issn:0251-0790.2003.03.037
2003-01-01
Abstract:Carbon fiber nanoelectrodes (200-300 nm) were firstly used to amperometricaly monitor the dopamine release from single PC12 cells with temporal resolution and especially more higher spatial resolution than those obtained by using microelectrodes. When the nanoelectrode was in a distance 1 mum above the PC12 cell, only one peak signal corresponding to a single vesicle exocytotic event was detected caused from stimulation with 1 mmol/L nicotine. The spatial difference of exocytosis was also detected by placing the electrode onto the different locations of the cell body, the results have demonstrated that the spatial distribution of dopamine in cells is not uniform and the time for stimulating secretion is very different. Nanoelectrodes electrochemical method can provide a powerful tool for observing the temporal and spatial characteristics of the secretion from single cells directly.
What problem does this paper attempt to address?