Absolute Calibration of Laser-induced Breakdown Spectroscopy Detection System by Using a Tungsten Halogen Lamp

Fei Teng,Pan Cong-yuan,Zeng Qiang,Du Xue-wei,Wang Qiu-ping
DOI: https://doi.org/10.3788/gzxb20184708.0847003
IF: 0.6
2018-01-01
ACTA PHOTONICA SINICA
Abstract:An absolute calibration setup for laser-induced breakdown spectroscopy detection system is built based on the radiometry theory. A tungsten halogen lamp equipped with an ultraviolet glass filter and a fused silica diffuser is used as standard light source for the calibration. Absolute calibration of a laser-induced breakdown spectroscopy detection system with a Czerny-Turner spectrometer in the ultraviolet range is performed. The absolute spectral response of the system in the range of 310-385 nm is obtained with an uncertainty less than 7% by 2 standard deviation estimate. The absolute spectral calibration provides a way to evaluate the hardware of the laser-induced breakdown spectroscopy detection system.