An Infrared Guided Patient Setup Method for Radiotherapy by Using a Modified Spirit Level

Peng Zhang,Jie Zhang,Guoping Shan,Dongxu Xu,Zhuoran Jiang,Weifeng Li,Jing Cai,Yun Ge,Ying Chen,Xiaolin Huang
DOI: https://doi.org/10.1166/jmihi.2018.2471
2018-01-01
Journal of Medical Imaging and Health Informatics
Abstract:To measure the tumor location error in six degrees of freedom (6DoFs) by using a combination of an infrared tracking technique and a modified spirit level, a new method is proposed. The method measures the tumor location error in 6DoFs by combing information from the computed tomography scans and the radiation treatment's coordinate system. Here, the radiation treatment's coordinate system is identified using the combination of a modified spirit level and an infrared tracking system. To evaluate the method, 8 patients with chest tumors were enrolled into the investigation and 63 data were collected. The difference of the proposal shifts that were given from a cone-beam computed tomography and the method was regarded as the estimation error of our method. The experiment showed that the method was capable of directing patient setup with translational tumor location errors of (2.78 mm, 3.23 mm, 2.30 mm) along left-right (LR), superior-inferior (SI) and anterior-posterior (AP) directions and rotational errors of (1.87 degrees, 2.06 degrees, 2.32 degrees) around the three axes. The method is a promising way to estimate tumor location error in 6DoFs. It may be also a feasible approach to avoid an inter-fraction error source that relates to the level degree of the treatment couch as much as possible.