Atomically resolved probe-type scanning tunnelling microscope for use in harsh vibrational cryogen-free superconducting magnet

Wenjie Meng,Jihao Wang,Yubin Hou,Mengqiao Sui,Haibiao Zhou,Junting Wang,Gang Wu,Jing Zhang,Fangchu Chen,Xuan Luo,Yupin Sun,Junyun Li,Qingyou Lu
DOI: https://doi.org/10.1016/j.ultramic.2019.06.006
IF: 2.994
2019-01-01
Ultramicroscopy
Abstract:•This is the first reported atomic resolution scanning tunneling microscope (STM) that is designed to work in a harsh vibrational cryogen-free superconducting magnet.•Large range of sample temperature, covering from 1.6 to 300 K.•Ultra-low noise down to 1.5 pA.Hz−1/2 when we set the current to 0.5nA and opened the feedback loop and drifting of 1.15 and 1.71 pm/min in X–Y plane and Z direction for the STM are achieved.•High quality STM images of graphite are obtained in sweeping the magnetic fields from 0–8 T.•Atomically resolved STM images of NbSe2 at 1.6 K are obtained, as well as scanning tunnelling spectrum (STS, i.e. dI/dV spectrum) data near its critical temperature, showing the formation process of the superconducting gap as a function of temperature.
What problem does this paper attempt to address?