Improvement of the Spatial Resolution of Laser-Induced Breakdown Spectroscopy with a Gated Signal Amplifier

Zhang Jianhua,Chen Yuqi,Li Runhua
DOI: https://doi.org/10.3788/cjl201744.0411003
2017-01-01
Chinese Journal of Lasers
Abstract:In order to realize high spatial resolution elemental microanalysis, agated signal amplifier is designed based on OPA 695operational amplifier and is applied in weak signal detection in orthogonal dual-pulse laser-ablation laser-induced breakdown spectroscopy. This gated amplifier rejected the influence of strong electronic Bremsstrahlung emission successfully, and the weak atomic emission could be selectively amplified, thus the spectral analysis sensitivity and spatial resolution could be improved.Major aluminum and minor copper elements in an aluminum standard sample are analyzed experimentally.Under current experimental condition, the spatial resolution has been reached 0.9and 1.2μm while analyzing aluminum and copper, respectively.These spatial resolutions are significantly improved in comparison with 2.9μm for aluminum and 6.2μm for copper obtained experimentally without using this gated signal amplifier.This gated amplifier is low cost and is helpful for high sensitive detection of weak signal accompanied with strong and time-resolvable background.
What problem does this paper attempt to address?