Method for enhancing stability in multi-beam microscopy

Yujia Huang,Yifan Wang,Cuifang Kuang,Xu Liu
DOI: https://doi.org/10.1088/0957-0233/27/10/105901
IF: 2.398
2016-01-01
Measurement Science and Technology
Abstract:A method based on close loop control of four degrees of freedom (4DF) is proposed to enhance angular and translational stability of beams in multi-beam microscopy including STED, RESOLFT and CARS, etc. Deviations of multi-beams can be measured and corrected by our module, which is composed of four degrees of freedom position sensitive detectors (4DF PSD) and two actuator mirrors (AM) with motor and piezo servos. An output crosslink matrix obtained by a self-learning process is used to control four actuators to compensate for 4DF independently in beam deviations. We realize a standard deviation within about 2 mu m at the entrance pupil plane (a spatial optical path of 180 cm for the whole system) using a compact stabilization system, which is equivalent to around 3 nm at the sample plane under the 100x objective lens with a focal length of 2 mm, corresponding to an improvement of stability by an order of magnitude. Our method can react fast in real time and compensate for large disturbances caused by air agitation or temperature variation.
What problem does this paper attempt to address?