Stereo Vision Sensor Field Calibration Method Based on Volume Template

陈刚,车仁生,叶东,黄庆成
DOI: https://doi.org/10.3321/j.issn:1004-924x.2004.06.014
2004-01-01
Optics and Precision Engineering
Abstract:A volume template-based method is proposed for the field calibration of the intrinsic and extrinsic parameters of stereo vision sensor, which adopts an ideal pin-hole model that ignores the nonlinear distortion of the cameras. Taking the transform matrix elements as unknown, when the group of 3-D target points coordinates and their image coordinates are known, they can be used in a linear algorithm to work out the unknown elements and the parameters needed. In the end, a volume calibration template is used to calibrate a stereo vision sensor, which measures a known length. The relative precision is up to 0.02%. Experimental results show that method needs no iteration and has a high operation speed, which needs to put the template only once for the field calibration, cameras' intrinsic and extrinsic parameters could be calibrated with high precision.
What problem does this paper attempt to address?