Calculation and measurement of effect of gold marker artifact on CyberKnife field dosimetry

QIAN Jin-dong,WANG Yun-lai,JU Zhong-jian,WANG Xiao-shen,ZHOU Yi-bing
DOI: https://doi.org/10.7687/J.ISSN.1003-8868.2014.11.092
2014-01-01
Abstract:Objective To analyze the influences of the gold marker artifact on Cyber Knife field dosimetry during CT scanning. Methods Solid water was scanned with CT simulator to obtain CT images with the gold marker at different depths.In the planning system, some single field radiation scheme was established with ionization chamber collector as the isocenter, and then the scheme was applied to the CT images with the specified artifact density for measurement. Results The gold marker had influences on the isocenter dose by 0.5%, and the artifact had influences on the dose and the offaxis absorbed dose at the same depth by 3%, which increased with the depth of the gold marker. When gifted with the fixed overlay for the artifact density, the artifact had influences on the dose by 1%. The gold marker artifact resulted in an apparent drop zone for the percentage depth dose(PDD) behind the marker, which could be weakened by the overlay of the artifact density. The marker made OCR curve flattened with few influences on it. Conclusion The overlay for the artifact decreases the effects of the gold marker artifact on the dose.