Development and Clinical Application of a New Biliary Teflon Stent

胡冰,周岱云,龚彪,王震,夏莉,王书智,孟建勤,王秀平
DOI: https://doi.org/10.3321/j.issn:0258-879x.2002.05.023
2002-01-01
Academic Journal of Second Military Medical University
Abstract:Objective:To develop a new type of polytetrafluoroethylene(Teflon) biliary stent for endoscopic use, and to evaluate its therapeutic effect.Methods:Eighty patients with malignant ultrahepatic bile duct obstruction were randomly divided into the test groups (with above stent) and control group (with 8.5F imported Teflon stent).The near and late effects were compared between the 2 groups. Results:Post procedure fever occurred in 22.2% patients in the test group and in 24.0% in the control group, hyperamylasia occurred in 27.8% and 27.3% of patients, and effective rate of jaundice control was 81.0% and 87.0% in patients in test and control group respectively.There was no significant difference in near effects between the 2 groups.The median stent patency was just the same (6 months) and median survival was 7 months in both group.Conclusion:The new stent has good capability and its main therapeutic parameters are similar to that of oversea products.
What problem does this paper attempt to address?