Dosimetric Verification of the Dose Distribution for a Cervix Intracavity Brachytherapy Using a Tow-dimension Ion Chamber Array

刘浩,王新,李公平,唐成琼,杜新香,穆塔力普江,马栋辉,王滢
DOI: https://doi.org/10.3969/j.issn.1005-202X.2014.04.004
2014-01-01
Abstract:objective: To verify the dose distribution for a cervix intracavity brachytherapy using a two-dimension ion chamber array in a phantom. Methods: Six cervix intercavity brachytherapy plans were selected to verify the dose distribution, the treatment length of the plans is 5 cm or 6 cm, the catheter of the plans is the intrauterine tube 15°,or the intrauterine tube 30°,or the intrauterine tube 45°. The catheters within the water-equivalent slab(SP34) were scanned with the computed tomography scanner(PHILIPS BIG BORE)and the images of the cathers was exported to the treatment system(TPS)(Oncetrol brachy 4.1).The catheters were marked in the TPS. The dwell position and the dwell time of the Ir-192 brachytherapy radiological source in the catheter were set up as the cervix brachytherapy plan and were exported to the afterloading treatment delivery console computers. The ion chamber arry was irradiated with the plan from the TPS. The thickness of the water equivalent slab between the ion array(IBA, MATRIXX) surface and the catheter is 1.6 cm. The data acquistion in MATRIXX was done in the movie mode.The treatment planning system dose distribution was calculated on the scan of the phantom and was exported to and compared in IBA Dosimetry OmniPro-I'mRT software. The quality of agreement was quantifyed by the gamma(γ) index(with3% delta dose and distance criterion 3 mm) for 6 sets of plans. Results: For one plan, compareing the measured and TPS caculated distributions,there is only one calculated dose plane agreement with 100% for γ≤1,the mean distance for 6 sets ofplans between the dose plane and the catheter plane is(2.3±0.1) cm。 Conclusions: The measured distribution is in agreement with the TPS-caculated distribution.It is feasible to verify relative dose distriutions for a Cervix Intracavity Brachytherapy.
What problem does this paper attempt to address?