Application of a Tow-dimension Ion Chamber Arrary for Quality Assurance in Brachytherapy

LIU Hao,LI Gong-ping,SUN Yan,TU Ni-sha,WU Heng,MA Dong-hui,WANG Ju-wu
DOI: https://doi.org/10.3969/j.issn.1005-202X.2011.04.004
2011-01-01
Abstract:Objective: To measure the dwell time of the radiological source during performing the brachytherapy plan,to compare the measured data with the designed data,and to investigate the reproducibility of reaching the dwell position of radiological source.Methods:We designed a brachytherapy plan: the treatment length was 15 cm,31 dwell positions were arranged at 0.5 cm intervals,the time weight of each position was equal,the total treatment time was 105 seconds.The transparent X-ray catheter for esophagus was used in the plan.We performed the plan ten times using the Kelinzhong afterloader system and measured the radiological source using the IBA MATRIXX in movie model with 0.1 second sample time.Results: The measured dwell time of each dwell position is 3.4 seconds,the total treatment time is 105.4 seconds and more 0.4% than the designed time.The relative deviation of main ion chamber data is less than 10%.Conclusions: It is feasible to verify the dwell time and the dwell position reproducibility of the radiological source in performing the treatment plan.The smaller the sample time is,the higher the precision of the dwell time is.