32P stent made by chemical electroplation to prev ent the restenosis after PTCA in rabbit iliac arteries

高东升,陈润芬,毛家亮,黄定九
DOI: https://doi.org/10.3969/j.issn.1001-1439.2002.07.015
2002-01-01
Journal of Clinical Cardiology
Abstract:Objective: To evaluate the effect of 32 P stent to prevent the restenosis after PTCA. Methods: Six 32 P β emitting stents made by chemical electroplation and paired with 6 nonradioactive stents were placed in double sides of 6 rabbits' iliac arteries. Animals were sacrificed after 30 days. Histological image analysis and angiography were performed. Result: There were no aneurism or thrombosis in the arteries with stents placed. Morphometric measurments demonstrated that the minimum lumen diameter (MLD) was greater within 32 P stents than nonradioactive stents (P 0.01 ). Furthermore, late loss, loss index and percent of restenosis in stent were lower in 32 P stents than those of control stents(P 0.05 ). The lumen area within 32 P stents were greater than within control stents, but the later had marked intimal area than the former (P 0.01 ). The edge effects were not observed in the present study. Conclusions: Low dose 32 P β emitting stent can prevent in stent restenosis significantly by inhibiting the intima growth and the edge effects may be avoided by using short balloon.