Chemiluminescence of diazotized luminol for selective determination of nitrite in water

GUO Yun-yun,ZHANG Sheng-hai,CUI Bei,WANG Li-jia,L(U) Jia-gen
DOI: https://doi.org/10.13595/j.cnki.issn1000-0720.2014.0123
2014-01-01
Abstract:A novel chemiluminescent method coupled with the flow injection technique for high selective determination of nitrite was proposed, based on the enhanced chemiluminescent reaction signal between diazotized luminol and hydrogen peroxide. The linear range for the concentration of nitrite was 4. 8 × 10-8~ 4. 8 × 10-6mol / L with good correlation coefficient of R2= 0. 9994 by the new method. The detection limit of the method was 1. 5 × 10-8mol / L. The relative standard deviations( RSD) for eleven replicates were 3. 6%. There is no significant difference between national standard method and the method for detecting the known concentration of nitrite. The proposed method has been applied to detection of NO2 –in tap water and boiling water. In conclusion,the described method has high sensitivity,good selectivity,and no secondary pollution.
What problem does this paper attempt to address?