Chemiluminescence detection of nitric oxide in seawater

FENG Wei-hua,LIU Chun-ying,YANG Gui-peng,LI Pei-feng,XUE Chao
2011-01-01
Acta Oceanologica Sinica
Abstract:A chemiluminescence method measuring the concentration of NO in seawater was established in this paper on the base of a reaction between the mixed solution of luminol-H2O2,the chemiluminescence probe,and NO,in which the chemiluminescence signal was emitted.The optimal results in the determination of NO concentration were under the conditions of a pH 9.2 buffer system with 1×10-2 mol/dm3 carbonate,8.4×10-5 mol/dm3 luminol,1×10-2 mol/dm3 H2O2,and under 20 ℃ and 1×10-3 mol/dm3 EDTA as a optimal masking agent.Under these optimal conditions,there was a significant linear relationship between the NO concentration and the peak area of chemiluminescence signal in the range of NO concentration from 1.4×10-12 to 9.8×10-10 mol/ dm3(R=0.997,P 0.000 1).The detection limit for NO was 1.4×10-12 mol/dm3(S/N).In 7 times of parallel determination to 1.2×10-11 mol/dm3NO aqueous solution,the relative standard deviation(RSD) was 13.6%.The NO concentration in the surface seawater of Jiaozhou Bay,China was successfully determined using this method and the results indicated that the concentrations of NO ranged from undetected to 7.1×10-11 mol/ dm3 with an average of(1.6±1.8)×10-11 mol/dm3.
What problem does this paper attempt to address?