Value of High Impedance Leads in Permanent Cardiac Pacing

马小川,王建清,张全发
DOI: https://doi.org/10.3969/j.issn.1004-583x.2004.13.007
2004-01-01
Abstract:Objective To evaluate the handling characteristics, the reliability and effect of high impedance leads in permanent cardiac pacing. Methods Six patients included as experiment group were implanted with high impedance atrial and ventricular leads. Other 24 patients were implanted with traditional steroid-eluting leads as control group. The original values of lead impedance and pacing thresholds were tested and compared by t-test between groups. The patients with high impedance leads were followed up at 1 week,1 month and 3 months postimplant. Lead impedance, pacing thresholds were observed during follow-up. Results The implanting procedure took ( 1.8± 0.5) hours. The original voltage thresholds had no significant variety(P 0.05),but the leads impedance had very significant variety between two groups(P 0.001).No significant variety was observed in pacing impedance or thresholds during follow-up. No dislocation of the leads occurred in the period of follow-up. No infections were observed. Conclusion The surgical handling of high impedance leads is no more difficult than traditional steroid-eluting leads. The pacing thresholds are satisfactory. The leads display higher impedance values and lower current values. Thus high impedance lead appears to result in a clinically relevant extension of generator longevity.