Study of the accuracy of dose delivery in brachytherapy

徐晓,常力方,杨天恩,陈丽
DOI: https://doi.org/10.3969/j.issn.1005-202X.2003.04.005
2003-01-01
Abstract:Purpose: To study the accuracy of dose delivery for brachytherapy in our Hospital. Method: Using apparent activity (Ci) in water method and Air Kerma Rate (cGy·m2·h-1) in-air method calibrated 192Ir source. The accuracy of the dose delivery was determined for HDR Nucletron Brachytherapy using a water box containing a PTW M23331 Ionization chamber in its center- 4 applicators were placed at 2 cm from the ionization chamber- equally spaced at 90° angles. Resules: In a comparison of sourcecalibration procedures for 192Ir HDR sources in water method found that the average deviations from the certificate's value was 4.38%- but the average deviation was 2.4% in-air method. The average deviation of the measured dose from the prescribed dose was 3.87%. Conclusions: This comparison showed that the accuracy of brachytherapy dose delivery has come up to the standard of ICRU 24# report.