Clinical Application of Biatrial Pacing Technique in Atrial Tach Yarrhythmias

王振东,郭涛,赵玲,王松丽,韩明华,肖践民,李淑敏,刘中梅,郑语今
DOI: https://doi.org/10.3969/j.issn.1001-1439.2002.07.001
2002-01-01
Journal of Clinical Cardiology
Abstract:Objective:To evaluate the efficacy and safety of biatrial pacing in treating paroxysmal atrial tachyarrhythmias in patients with interatrial conduction block. Method:Under fluoroscopy in frontal and left anterior oblique projection, three leads were placed into coronary sinus (CS), right ventricle (RV) and right atrium (RA) respectively via left subclavian venous route in fifteen patients. A new bi polar lead were composed of CS lead and RA lead through a Y shape adapter after threshold tests. AAT, DDD or DDTA pacing mode was applied after implantation, respectively.Result:One patient abandoned the therapy because of dislocation of CS lead in operation. CS leads dislocated in two patients after operation and both were replaced successfully. Two patients experienced pacemaker mediated tachycardias because of programming inappropriately. The other patients had no complications. There were obvious efficacy in nine patients and partial efficacy in two patients while three of them have no efficacy during follow up (range two months thirty two months).Conclusion:Biatrial pacing can effectively prevent and treat paroxysmal atrial flutter and fibrillation in patients with advanced interatrial conduction block. The ordinary ventricle lead, which is safe, cheap, effective and low in dislocation rate for left atrial pacing through coronary sinus, deserves to be generalized and applied.