Registration accuracy of neuronavigation: laboratorial study

杜固宏,周良辅,毛颖
DOI: https://doi.org/10.3969/j.issn.1009-122X.2002.04.008
2002-01-01
Abstract:Objective To investigate the influence factors of neuronavigation registration accuracy for providing theoretical guidance of clinical application.Method A neuronavigation phantom was designed and produced.The phantom was imaged by CT and MRI,and then the imaging data were transferred to the navigation system.The registration errors(MFE and PA)were compared among the study groups that the fiducial number and its arrangement was different from each other.Result In CT-based procedures,a significant difference of MFE was found between the four -fiducial group and the others.A significant difference of PA was found among all groups.In MRI -based procedures,a significant difference of MFE was found among all groups.No significant difference of PA was found among the groups except the four -fiducial group and the six-fiducial group.Registration accuracy was significantly higher in the groups with the fiducials arranged in a scattered pattern than that with the fiducials arranged in a linear pattern.Conclusion Registration error decreased with increase of the fiducial number.In order to obtain a high accuracy,we recommended that at least6fiducials should be used for registration of the CT-based neuronavigation and10fiducials for the MRI -based neuronavigation,meanwhile,the fiducials should be arranged in a scattered pattern.