Study on monitoring of glucose concentration with optical coherence tomography

Dong-ming LI,Shu-hai JIA
DOI: https://doi.org/10.16136/j.joel.2014.11.032
2014-01-01
Abstract:Optical coherence tomography(OCT)is one of the rapidly developed techniques for bio-tomography in the last few decades.It can acquire the high resolution cross-section image of micro mechanism in the interior organism tissue.Due to its advantages,including rapid imaging,non-invasive detecting and high resolution,it has been widely applied in biomedical fields,especially in early diagnosis and dynamic monitoring for the systems of eyes,skin,oral cavity and cardiovascular.The Michelson interferometer for optical coherence tomography technology is uesed,an OCT system is designed and combined.The physical level concentrations of glucose solutions are measured.The relationship between the distance of the interference wave images on the highest and the lowest and the sample concentration is explored.In the laboratory,the samples on physiological level are obtained from the compound 5% glucose solution.The results show that the linear coefficient is 0.999 8,and the standard deviation is 0.0039.Using OCT system by the related features between sample concentration and the distance of interference image wave from the highest to the lowest can measure the concentration of glucose solution accurately and rapidly.This new method can be used for the study of non-invasive blood glucose measurement instrument.
What problem does this paper attempt to address?