Precision of Glucose Measurement in Intralipid Suspensions with Optical Coherence Tomography

Shu Feng,Kehong Yuan,Datian Ye
DOI: https://doi.org/10.1109/itab.2008.4570660
2008-01-01
Abstract:Optical coherence tomography (OCT) presents its capability of noninvasive glucose measurement in tissue-simulating phantoms and biological tissues. However, speckle noise substantially deteriorates the accuracy of the measurements with this technique. The speckle noise can be suppressed by averaging of in-depth scans. We studied the suppression of speckle noise for accurate measurement of backscattering signal and the maximum precision of glucose measurement in Intralipid suspensions with the OCT technique. Our results show that the precision of glucose measurement can achieve +/- 4.4 mM for 10% Intralipid and +/- 2.2 mM for 3% Intralipid after the averaging of 50,000 scans.
What problem does this paper attempt to address?