Heat pain threshold measurement : repeatability and sensitivity

王宁华,许云影
DOI: https://doi.org/10.3969/j.issn.1001-1242.2003.01.005
2003-01-01
Abstract:Objective: To systematically define the sensitivity and repeatability of heat pain threshold, using the Medoc TSA-2001 device. Method: The study on repeatability was performed on 8 young normal subjects on the knee, foot, hand and trunk, using the Method of Limits. Tests were repeated one and two weeks following the first one. In the study on sensitivity, 10 young healthy subjects performed subjective psychophysical estimates involving warm and hot cutaneous stimuli. Ten levels of temperature ranging from 32℃ to 50℃ were applied for 10 seconds to the left knee in a random order. Subjects gave continuous ratings of perceived temperature and pain intensity using the visual analogue scale (VAS). Intraclass correlation coefficient (ICC) and linear regression were used for data analysis in repeatability and sensitivity respectively. Result: ICC of the tested sites ranged from 0.87 to 0.95. There was a linear correlation between heat stimulus intensity and subjective thermal sensation, with a high value of r =0.93 and a slope of 0.54. Conclusion: Heat pain threshold measurements were not only highly repeatable but also highly sensitive. Under the present experimental paradigm, heat pain processing by the nervous system appears to be a linear function of the stimulus intensity in young humans.
What problem does this paper attempt to address?