Measurement and Analysis of Thermal Physical Properties in Deep Tissue

夏全龙,夏雅琴,南群,彭见曙
DOI: https://doi.org/10.3969/j.issn.1002-3208.2003.01.011
2003-01-01
Abstract:A measurement method of thermal properties and blood perfusion rate of organic body tissue is described in this paper. Thermistor probe was inserted into local tissue of interest, and electric power was input to raise initial temperature to desired mean temperature. The required electrical power for such purpose depends on the heat transfer and blood perfusion characteristics. A mathematical model for coupling thermistor and measurement medium was established. Then, thermal conductivity, thermal diffusivity and perfusion could be calculated by using this model. The conductivity of liquid medium and dead medium was measured first, which was used to verify the accuracy of the measurement system. Then the conductivity and perfusion rate of the fluid medium was measured. The experimental results of fluid and dead medium complied with what is stated literature, and the results of fluid were consistent with our analysis.
What problem does this paper attempt to address?