Accurate Attenuation Correction for Self-collimating SPECT Imaging
J. Wu,L. Guo,H. Liu,Z. Lyu,Y. Zhao,D. Zhang,Z.-X. He,Y. Liu,T. Ma
DOI: https://doi.org/10.1109/nssmicrtsd49126.2023.10337852
2023-01-01
Abstract:Recently, our group proposed a self-collimating SPECT system with multi-layer interspaced mosaic detectors to achieve high resolution and sensitivity simultaneously. In this study, we investigated the effect of attenuation on image quality and applied an accurate attenuation correction during reconstruction for self-collimating SPECT based on a half-ring and a full-ring system, respectively. The attenuation factor was calculated using a fast ray-tracing algorithm. The attenuation correction was performed during OSEM reconstruction by multiplying the attenuation factor by the system matrix. Noise-free and noisy analytical simulations were performed using an XCAT phantom with 1 slice. GATE simulations were performed using an XCAT phantom with 20 slices. The reconstructed image quality was evaluated quantitatively using the absolute relative bias between the reconstructed images and the phantom image. The simulation results showed that attenuation correction greatly improved the image quality qualitatively and quantitatively. The results also showed that the full-ring system was preferable than the half-ring system to avoid attenuation-caused artifact when statistical noise was introduced. The GATE simulation results of the full-ring system showed that the absolute quantification biases were reduced after attenuation correction from 82.0% to 4.2%, from 87.7% to 7.7%, from 24.4% to 2.2%, from 95.0% to 9.5% for 4 ROIs, respectively. In conclusion, attenuation correction is crucial for accurate quantification of the self-collimating SPECT images and the full-ring system is preferable than the half-ring system when both statistical noise and attenuation effect exist in practical applications.