Image Reconstruction in 3D Short-Scan SPECT

Xiaochuan Pan,Chien-Min Kao,Charles Metz,Alexander Kiselev
2001-01-01
Abstract:Physical factors such as photon attenuation degrade image quality and quantitative accuracy in single-photon emission computed tomography (SPECT). It is often considered (especially in situations with non-uniform attenuation and distance-dependent spatial resolution (DDSR)) that adequate compensation for the effects of these physical factors requires data acquired over 2 . However, as the analysis in this work suggests, one may need data acquired only over to correct adequately for the effects of some of the physical factors. Reduction of the scanning angle in SPECT imaging is desirable because it can reduce the scanning time and thus minimize patient-motion and other artifacts, and because scans over less than 2 can allow the detector to be rotated at a fix ed distance closer to the patient. Also, in certain cases (e.g., in cardiac SPECT), one can choose the scanning angular range for obtaining maximum numbers of photons. This work focuses on investigation of accurate image reconstruction in 3D SPECT from data that are acquired with parallel-beam collimation only over and that contain the effects of photon attenuation (either uniform or non-uniform) and DDSR. For simplicity, we refer to such a scanning configuration as 3D short-scan SPECT. This work may have significant theoretical as well as practical implications for image reconstruction in SPECT.