Clinical Investigation on Determination of Projection Angle of Each Image in Rotational Digital Subtraction Cerebral Angiography

孙立军,张学昕,白洪涛,贺洪德,余厚军,牟轩沁,李敬邦
DOI: https://doi.org/10.3969/j.issn.1002-1671.2004.03.001
2004-01-01
Abstract:Objective To study the determination of projection angle for each image in rotational digital subtraction cerebral angiography(RDSCA).Methods Several rotational data were obtained using a Philips V5000 digital imaging system by phantom experiment, and a mathematic expression was produced which could describes approximately running rule of the angiographic unit,and then each projection angle of RDSCA in 23 patients with cerebral arterial abnormalities was calculated by a programmable handset.Results Projection angle of each interested image was calculated immediately by the handset in all RDSCAs of 23 cases with intracranial aneurysm,arteriovenous malformation and intracranial meningioma. Among of these, 14 intracranial aneurysms, and 1 left fossa meningioma and 8 cerebral arteriovenous malformations were to be found. Compared with optimal image in rotational images, the image by calculation was resemblance in showing origins,running,distribtions and contours of diseased segments of arteries and afferent arteries as well as draining veins.Conclusion Determination in real time of each image in RDSCA is of significance for accurate diagnosis and intervention of intracranial vascular abnormalities.