Study on Measurement of Dielectric Constant of ER Fluids Granule Phase

唐宏,杨俊刚,尹剑波,赵晓鹏
DOI: https://doi.org/10.3969/j.issn.1001-2028.2003.01.002
2003-01-01
Abstract:Based on the mingling rule of dielectric constant of two-phase-even-mixed compound and Wheatstone bridge method of dielectric constant measurement, two methods have been developed. They are air-subtracted method and oil-mixed method for determining the dielectric constant of fine powder used in ER fluids as dispersant. The air-subtracted method avoids the influence of air void produced in the ordinary plate-pressed method; therefore increases precision. With the methods, several kinds of materials have been tested and satisfactory results have been achieved. In order to improve mechanic properties, the dielectric constant and conductivity of rare earth element doped TiO2 were tested and it was find that the shear stress of ER fluid increases by approximately 6 times at 0.1 RE/TiO2 mole-rate.
What problem does this paper attempt to address?