Comparison of Sensitivity of Different Testing Methods in Assessing the Cytotoxicity of Bioprosthetic Heart Valve Materials

程述森,石应康,梁卫东,李炜如,熊英
DOI: https://doi.org/10.3969/j.issn.1002-0179.1999.02.015
1999-01-01
Abstract:Objective:The difference in cytotoxicity of bioprosthetic cardiac valve materials was compared by using agar overlay and direct contact assay methods.Method:The material was crosslinked by glutaral dehyde and stored in 4% formaldehyde.The cells used for testing were human embryonic fibroblasts and mouse L-929 cells.The agar overlay and direct contact assay methods were used and their results were analyzed and compared.The cell proliferative inhibition index(CPII)was calculated for the testing materials being at the periods before,10,20 and 30 days after immerising the testing materials in the cellular suspension fluid.Conclusion:It was concluded that the direct contact method was more sensitive in the detection of cytotoxicity of the material than that of the agar overlay method.It was considered that in testing the cytotoxicity of the prosthetic heart valve material the direct contact method was preferable because it was more sensitive and accurate.
What problem does this paper attempt to address?