Abstract:Patients with injury of anterior cruciate ligament need to be repaired by the reconstruction surgery in time,and one of the keys to the surgery is the tunnel placement of implant. For the problem of implant tunnel placement depending on higher requirements of medical technology,personalized navigation template based on reverse engineering and rapid prototyping technology was used as aids to treat double-bundle reconstruction of anterior cruciate ligament,and personalized surgical plans were made for patients. In addition,the positioning accuracy of navigation template and operation quality were evaluated by combining the positions of the virtual preoperative planning,the simulated surgery based 1∶ 1 prosthesis and the postoperative clinical tunnel together. The results show that,in the clinical postoperative measurement,the center of anterior-medial bundle of tibia accounts for maximum distance of the tibia platform between the front and back edge,inside and out-side edge 34. 61% and 45. 87%, within the scope of reference data 37. 6% ± 3. 6% and37. 6% ± 3. 2%; In the same way,the center of posterolateral bundle of tibia accounts for 46. 41%and 50. 28%,within the scope of reference data 50. 1% ± 5. 0% and 51. 2% ± 2. 4%,too. To sum up,this method can simplify the operation,and the location accuracy of the navigation template is high.