Determination of Diclofenac's Concentration in Plasma by UPLC-MS/MS

GU Ji-hong,DONG Ji,MIAO Li-yan
DOI: https://doi.org/10.14066/j.cnki.cn21-1349/r.2014.12.009
2014-01-01
Abstract:Objective To establish a UPLC-M S/M S method for the determination of diclofenac in human plasma.Methods The plasma procedure involved a single-step protein precipitation by acetonitrile,then a chromatography was proceeded on a Waters ACQUITY UPLC BEH C18column(50 mm × 2.1 mm,1.7 μm) using a mobile phase consisted of acetonitrile-0.06%(φ) formic acid(V ∶ V = 40∶ 60) at a flow rate of 0.7 m L·min-1.The protonated ions of analytes were detected in positive ionization in multiple reaction monitoring mode(M RM).The mass transition pairs of m/z 296.03 →249.98 and 358.10 →138.92 were used to detect diclofenac and indomethacin.Results The calibration curve was linear over a concentration range of 10 to 3 000 μg·L-1(r = 0.998 0) and the low limit of quantitation was 10 μg·L-1.The extraction recovery rates were ranged from 95.1% to 101.3%,with-day RSD and between-day RSD were all less than 10%.Conclusions The method is simple and quick,with high specificity and sensitivity,and suitable for the research of diclofenac pharmacokinetics.
What problem does this paper attempt to address?