Methods of Pulse-Blood Oxygen Sensor's Noise Removal

陈涛,毛文林,叶继伦
DOI: https://doi.org/10.3969/j.issn.1000-9787.2004.03.021
2004-01-01
Abstract:Under the condition of low perfusion, the measurement of non-invasive saturation of pulse-blood oxygen is low-accuracy, lower signal-to-noise ratio(SNR) is one of the crucial causations. Removing the pulse-blood oxygen sensor's noise is the key way to improve the accuracy of oximeter's measurement. The testing data show that: after improving the sensor from three aspects, that is, ambient light noise, electro-magnetic Shielding and motion artifacts removal, the sensor's noise level is reduced accordingly in different degrees, and the oximeter's performance under the condition of low perfusion is ameliorated markedly, which proved powerfully that the method of noise removal is efficient.