Determination of Nisin Activity by Oxford Plate Assay System

伊守亮,肖林,顾正华,章克昌
DOI: https://doi.org/10.3321/j.issn:1673-1689.2004.04.011
2004-01-01
Abstract:The conditions of detecting Nisin activities by oxford plate assay system were studied in this paper. A couple of effecting factors were investigated. The optimal detecting conditions were as follows: monolayer medium loading quantity of 15 mL, Micrococcus luteus suspension concentration of 10~(9) CFU/mL, the loading quantity in Oxford Plate of 100 μL, 1 g/dL agar concentration, 1 g/dL Na_(2)HPO_(4)·12H_2O concentration, and the medium pH of 7.0. Under these conditions, a better linear relation was obtained when the Nisin concentration ranging from 5 IU/mL to 100 IU/mL.