Validation of SenseWear in Estimating Energy Expenditure During Exercise in High Temperature and Humidity Environment

Jin-tao QU,Fan-xing ZENG,Wen-ping FENG
DOI: https://doi.org/10.19582/j.cnki.11-3785/g8.2015.06.012
2015-01-01
Abstract:Objective: This study investigated the validation of Sense Wear( SWA) in estimating energy expenditure during exercise in high temperature and humidity environment for application of SWA,by using IC method as a reference. Methods: Ten male adults underwent 20 min treadmill exercise at 40% VO2max,60% VO2max,80%VO2maxrespectively in high temperature and humidity environment( temperature( 33 ± 1) ℃,relative humidity( 65 ± 3) %). Energy expenditure was measured by fixed gas metabolism apparatus( Cortex Meta Lyzer 3B) and SWA( Sense Wear Pro3 Armband). Data was analyzed by the methods of paired t test,regression analysis,correlation analysis,ICC coefficient and Bland-Altman. Results: SWA underestimated the energy expenditure of quiet state( 7. 6%),60% VO2maxintensity exercise( 15. 4%) and 80% VO2maxintensity exercise( 28. 8%),but it overestimated the energy expenditure of 40% VO2maxintensity exercise( 15. 6%). The relevance and consistency of SWA and IC in predicting energy consumption were good( r = 0. 73- 0. 85,IC = 0. 77- 0. 90). Conclusion: In high temperature and humidity environment,there are high relevance and high consistency between the predicting energy expenditure of SWA and IC. SWA is more suitable for estimating energy expenditure in low intensity exercise.