Determination the Optimal Monochromatic Energy and ASIR Blending Level in Coronary CTA with Spectral CT Imaging

李勇,孔祥泉,王成伟,宋娟,杨帆,郭宁,沈云
DOI: https://doi.org/10.13437/j.cnki.jcr.2015.06.032
2015-01-01
Abstract:Objective To investigate the optimal monochromatic energy and ASIR blending level for coronary CTA with spectral CT imaging. Methods 30 consecutive patients with heart rate ≤65 beats / min underwent prospective ECG-gated coronary CTA with dual energy spectral mode on a 64-slice CT scanner( Discovery CT750 HD,GE Healthcare). 5 groups of monochromatic images( 60 ke V,65 ke V,70 ke V,75 ke V and 80 ke V),each with( 0%,20%,40%,60%,80% and100%) six ASIR blending level were reconstructed. Polychromatic images( 140 k Vp) were also reconstructed. Signal-tonoise ratio( SNR),contrast-to-noise ratio( CNR),and image quality( with a 5-point Likert scale) were assessed in all above monochromatic reconstructions and compared with the polychromatic image. Results Under each monochromatic level,image noise gradually decreased as the ASIR propotion increase( each P 0. 001). Compared with polychromatic images,reconstructions with 60 ke V and 100% ASIR showed the highest improvement in contrast-to-noise ratio( 144%; P 0. 001) and signal-to-noise( 173%; P 0. 001). Image quality reached a plateau at 65- 75 ke V with 40%- 60%ASIR blending,at which,quality score was 40% higher than polychromatic images( P 0. 001). Conclusion In coronary CTA with spectral CT imaging,monochromatic energy( 60- 65 ke V) with( 40%- 60%) ASIR blending level provide significant image quality improvement.